M210A1

Amplifier

The M210А1 Amplifier provides:

  • verification and calibration of low-power voltage transformers (LPVT) as part of any VT calibration system;
  • conversion of any secondary voltage of a traditional VT into a value equal to the nominal voltage of the reference VT.

FEATURES AND BENEFITS

  • Continuous range of rated secondary voltages of verified LPVT
  • Continuous range of rated secondary voltages of verified traditional VT
  • High conversion accuracy
  • Simple use by intuitive user interface
  • Storage of individual voltage settings
  • Possibility of using with automated VT calibration system

 


APPLICATIONS

M210A1 is used by:

  • Manufacturers of LPVT and VT
  • Calibration Laboratories
  • Metrology Institutes
Input 1:
Rated input voltage range (Uin1r)0,1 V … 4 V
Input voltage resolution0,001 V
Additional factors× 1/√3; × 1/3
Rated operating frequency50/60 Hz
Input impedance, represented as a parallel resistor-capacitor circuit2 МΩ ±5% || C ≤ 50 pF
Input 2:
Rated input voltage range (Uin2r)30 V … 245 V
Input voltage resolution0,1 V
Additional factors× 1/√3; × 1/3
Rated operating frequency50/60 Hz
Input impedance, represented as a parallel resistor-capacitor circuit2 МΩ ± 5% || C ≤ 50 pF
Output:
Rated output voltage range (Uoutr)30 V … 100 V
Output voltage range0 … 120 V
Output voltage resolution0,1 V
Additional factors× 1/√3; × 1/3
External burdenR ≥ 500 kΩ; C ≤ 5 nF
Accuracy:
Ratio error (Uoutr/Uinr)
0,02·Uoutr ≤ Uout < 0,2·Uoutr
0,2·Uoutr ≤ Uout < 0,8·Uoutr
0,8·Uoutr ≤ Uout < 1,9·Uoutr

± 0,1 %
± 0,04 %
± 0,02 %
Phase displacement
0,02·Uoutr ≤ Uout < 0,2·Uoutr
0,2·Uoutr ≤ Uout < 0,8·Uoutr
0,8·Uoutr ≤ Uout < 1,9·Uoutr

± 5 min (± 0,15 crad)
± 2 min (± 0,06 crad)
± 1 min (± 0,03 crad)
Power mains:
Voltage
Rated frequency

100…240 V
50/60 Hz
Operating temperature
Relative humidity
0...40 °C
up to 80 % non-condensing
Size
Weight
245 × 135 × 325 mm
5 kg

CL – low-voltage measuring capacitor; CH – high-voltage measuring capacitor;
CC – current comparator with automatic balancing.

Fig.1 Test circuit for LPVT by use the Voltage Transformer Test Set M210/M211

VTN – standard transformer
Fig.2 Test circuit for LPVT by use the Comparator M100

VTN – standard transformer; VTX– transformer under calibration
Fig.3 Test circuit for traditional voltage transformers with different rated secondary voltages