M100A1

V-I Converter

V-I Converter M100A1 is used to expand the functionality of the Comparator M100 and other traditional comparison devices during the calibration and verification of low power current transformers (LPCT) with an analog output in the form of an alternating voltage.
M100A1 converts secondary voltage of LPCT into current

    FEATURES AND BENEFITS

    • Continuous range of rated secondary voltages of verified LPCT
    • Calibration of LPCT of class 0.1 and less accurate
    • Ability to use with any comparison device
    • Simple use by intuitive user interface
    • Storage of individual voltage settings, which can be corrected as required
    • Possibility of using with automated CT calibration system

     


    APPLICATIONS

    M100A1 is used by:

    • Manufacturers of LPCT
    • Metrology Institutes
    • Calibration laboratories
    • On-site testing of LPCT

    Name

    Value

    Rated output current

    1 A

    Output current range

    0.01 … 2 A (1…200%)

    LPCT rated secondary voltages range ( Usr)

    0.02 … 4.00 V

    Step of changing the LPCT rated secondary voltage

    0.1 mV (0.02 ≤ Usr < 0.1 V)
    1 mV (0.1 ≤ Usr < 1 V)
    10 mV (1 ≤ Usr ≤ 4 V)

    Input impedance

    2 MOhm || 50 pF

    Rated frequency

    50 or 60 Hz

    Output burden

    ≤ 0,2 Ohm

    The M100A1 errors:

    Output current

    0.01 A (1%)

    0.05 A (5%)

    0.2…2 A (20…200 %)

    Ratio error εU-I

    ±0.1%

    ±0.03%

    ±0.015%

    Phase displacement ΔφU-I

    ±3 min

    ±1.5 min

    ±0.50 min

    Phase displacement ΔφU-I

    ±0.09 crad

    ±0.05 crad

    ±0.015 crad

    Power mains
    Voltage
    Rated frequency

    100…240 V
    50/60 Hz

    Operating temperature

    0...40 °C

    Relative humidity

    up to 80 % non-condensing

    Size

    245 × 135 × 325 mm

    Weight

    5 kg

    LPCT testing