Electronic Current Burden
Electronic Current Burden M402 is used as a load in the secondary circuits of Current Transformers during their verification and calibration
With the wide programmable burden power range provided, most national and international standard burdens or specific customer values can be set
FEATURES AND BENEFITS
- Wide range of burden values according to standards: IEC 61869-2, IEC 60044-1, IEEE C57.13, VDE 0414-1, GOST 23624, GOST 7746
- Standard and non-standard burden power values at rated secondary currents: 1; 2; 5 А
- Rated burden power range:
0; 0,5…100 V·А, cosφ = 0,5…1 at 50 and 60 Hz
- Cable, contact and internal test set (Comparator) resistances are compensated by four-conductor connection
- Simple use by intuitive user interface
- Storage of individual burden settings, which can be retrieved as required
- Possibility of using with automated CT calibration systems
- The burden values range can be extented by connecting several devices in series
- Small size and low weight
APPLICATIONS
M402 is used by:
- Manufacturers of Current Transformers
- On-site testing of Current Transformers
- Calibration Laboratories
- Metrology Institutes
Rated current | 1; 2; 5 А |
Current frequency | 50; 60 Hz |
Current range | 1...200 % |
Range of burden power values | 0; 0,5...100 V·A |
Step of changing the burden value | 0,01 V·A |
Range of cosφ (power factor) | 0,5...1 |
Step of changing cosφ (power factor) | 0,01 |
Active and reactive power error | ± 0,03·S* |
Power mains Rated voltage Rated frequency | 110...240 V |
Operating temperature | 0...40 ºС |
Relative humidity | up to 95 % non-condensing |
Size | 460 × 145 × 465 mm |
Weight | 15,5 kg |
* Set burden power value |